GenPure™ Ultrapure Water System with UV-Photo-Oxidation and Ultrafiltration Membrane, Thermo Fisher Scientific
Manufacturer:
Thermo Fisher Scientific
SKU: 21-8990-11
or call us at 1-800-727-0467
Thermo Scientific™ Barnstead™ GenPure™ Water Purification Systems
Produce Type 1 ultrapure water with consistent quality for your most demanding and sensitive applications with the Thermo Scientific™ Barnstead™ GenPure™ water purification system. This system offers outstanding control, high-precision and USP-complying conductivity measurement with temperature compensation that can be switched on or off. All GenPure™ systems exceed international standard ASTM Type I, ISO 3696 Grade 1, ASTM D1193 and CLSI-CLRW.
- Drop-by-drop to full 2 L/min dispensing for controlled dispensing
Feed Water Inlet: 3/4in NPT
Feed Water Type: Municipal Water Pretreated by Reverse Osmosis, Ion Exchange, or Distillation
Other products by Thermo Fisher Scientific
| Manufacturer | Thermo Fisher Scientific |
|---|---|
| Promo Text | N/A |
All GenPure systems exceed international standard ASTM Type I, ISO 3696 Grade 1, ASTM D1193 and CLSI-CLRW
Integrated feed water monitoring
- An additional measuring cell monitors feed water conductivity. Any exceeding of the limiting value is immediately displayed maximizing cartridge life
-
Online TOC measurement with uV intensity monitoring
- Real-time TOC monitoring for continual measurement of organic substances present in the water
- Continuous monitoring of UV intensity removes the risk of false TOC values due to diminished UV intensity
- Fault message alerts user when the UV lamp intensity is no longer sufficient for accurate TOC measurements
- UV Photo-oxidation 185/254nm
- High performance UV assembly reliably reduces the content of microorganisms and their metabolites
Ready-to-use
- Feed water pressure switch, wall bracket, UV lamp, ultrapure cartridge, sterile filter, all included with one part number
Measurement of conductivity/resistivity
- Conductivity cells are carefully calibrated prior to each measurement via built-in reference resistance with cell constants at 0.01cm-1
- Temperature measurements are made by a platinum chip sensor with ± 0.1°C accuracy
Quick-connect cartridge replacement
- Aquastop quick-connect capability enables cartridge replacement within seconds even mid-operation
-
GLP-compliant documentation
- Real-time clock and code-protected operating system prevents unauthorized changes to system settings
- RS-232 interface with adjustable send - interval for safe data transfer of all measured data, faults, date and time to a PC computer or log printer
- Digital microprocessor control automatically monitors and stores faults from the last four weeks
- USP-compliant conductivity measurement with temperature compensation can be switched on or off
Tilting control panels for easy viewing
- The ergonomically designed controller tilts for optimal key pressing and easy readability
- Illuminated four-line alpha-numeric display