No more waiting for pH readings to stabilize when measuring samples of varying temperatures. Our Thermo Scientific™ Orion™ ROSS Ultra™ Triode™ pH electrodes combine the stability, precision and speed of response of the ROSS Ultra™ line with integrated temperature measurement. Orion™ ROSS electrodes contain no Silver or Silver Chloride and incorporate a double-junction design, which protects the sample from contamination and makes the electrodes compatible with TRIS buffers.
Other products by Thermo Fisher Scientific
| Manufacturer | Thermo Fisher Scientific |
|---|---|
| Promo Text | N/A |
- Material: Chemical resistant glass body
- 8 pin miniDIN connector for Orion Star meters
- pH range: 0 to 14
- Temperature range: 0 to 100°C
- Internal reference: ROSS
- Glass fiber junction
- Fill solution: 3M KCl, Orion 810007
- Dimension: 120 x 12mm (L x Dia.)
- pH precision: 0.01
- pH accuracy: 0.03pH for buffers using automatic temperature compensation
- Temperature accuracy: 1.0°C
- Isopotential point: 7
- Double junction design prevents sample contamination
- Filling solution can be changed when potassium or chloride in the sample is not desired
- Electrode drift is less than 0.002pH per day, so restandardization is minimized
- Stable to 0.01pH within 30seconds in 6.86 between 25°C and 75°C
Storage Conditions: Short Term Storage (up to one week)
Soak the electrode in ROSS* pH electrode storage solution, Cat. No. 810001, orstandard pH electrode storage solution, Cat. No. 910001. For enhanced electrodeperformance, use ROSS pH electrode storage solution.
Long Term Storage (more than one week)
Fill the reference chamber and securely cover the filling hole. Cover thesensing element and reference junction with the protective cap, sleeve orstorage bottle containing a few drops of storage solution. Before returning theelectrode to use, prepare it as a new electrode.
Warranty and Service Offering: 24 months